inkubator in30
Incubator IN30
Standard delivery Incubator
The heating of this incubator is optimally tuned for both natural convection and forced air circulation; the fan can also be switched off completely, and valuable chamber loads for research, pharmaceutics, medicine and food chemistry are warmed up very carefully.
On this page, you can find all the essential technical data on our incubator. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@.
Sketch Incubator IN30
Properties
Control of standard components
Temperature 1 Pt100 sensor DIN class A in 4-wire-circuit
ControlCOCKPIT adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Temperature
Set temperature range in C min. 5C above ambient up to +80C
resolution of display for setpoint and actual temperature values 0.1C
Control technology
Function SetpointWAIT the process time does not start until the set temperature is reached
Language setting German/English/Spanish/French
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Ventilation
natural convection
Fresh air admixture adjustment of pre-heated fresh air admixture by air flap control in 10 % steps
Vent vent connection with restrictor flap
Communication
Documentation programme stored in case of power failure
Programming AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
Safety
Temperature control adjustable electronic overtemperature monitor and mechanical temperature limit